Please use this identifier to cite or link to this item: https://dipositint.ub.edu/dspace/handle/2445/139347
Title: Ex-situ antimony screen-printed carbon electrode for voltammetric determination of Ni(II) - ions in wastewater
Author: Barceló, Cristian
Serrano i Plana, Núria
Ariño Blasco, Cristina
Díaz Cruz, José Manuel
Esteban i Cortada, Miquel
Keywords: Níquel
Voltametria
Anàlisi electroquímica
Nickel
Voltammetry
Electrochemical analysis
Issue Date: 28-Oct-2015
Publisher: Wiley-VCH
Abstract: An Ex-situ antimony film screen-printed carbon electrode (Ex-situ SbSPCE) was successfully applied for the determination of Ni(II), by means of adsorptive stripping voltammetry using dimethylglyoxime as complexing agent, in a certified reference wastewater sample. This electrode is proposed as an alternative to more conventional antimony film electrodes. Ex-situ SbSPCE was analytically characterized and the obtained parameters suggest that Ex-situ SbSPCE behaves much better than both Ex-situ BiSPCE and BispSPE for Ni(II) determination. The results confirm the applicability of Ex-situ SbSPCE for the determination of low concentration levels of Ni(II) in natural samples with a very high reproducibility and good trueness.
Note: Versió postprint del document publicat a: https://doi.org/10.1002/elan.201500511
It is part of: Electroanalysis, 2015, vol. 28, num. 3, p. 640-644
URI: http://hdl.handle.net/2445/139347
Related resource: https://doi.org/10.1002/elan.201500511
ISSN: 1040-0397
Appears in Collections:Articles publicats en revistes (Enginyeria Química i Química Analítica)